CSpace  > 精准医疗单分子诊断技术研究中心
A microfluidic chip for terahertz spectral detection
Liu, Mengwan1,2; Zhou, Daming2; Zhang, Mingkun2; Cui, Hong-Liang1,2; Wang, Deqiang2
2016
摘要Fluidic chips as sample storage devices have been widely used to control and analyze traces of solutions by terahertz time domain spectroscopy (THz-TDS). This paper presents a fluidic chip fabricated with ultra-thick micro-fluidic structures utilizing SU-8 photoresist for THz spectral detection. The chip consists of 2 pieces of quartz windows and a photoresist structure serves as liquid cell between the windows. In order to reduce the attenuation of THz energy by the windows and guarantee measurement stability, an infrared quartz with thickness of 1mm is adopted as the windows. For the purpose of improving the signal-to-noise ratio (SNR) in the frequency band between 1.0 THz-1.5 THz, and making the liquid film thickness match the THz optical path, the structure depth is designed to be 50μm. Water absorption coefficient at 0.1THz-1.5THz was detected with this chip. The result indicates that the chip has a good stability for terahertz spectral detection. © 2016 IEEE.
语种英语
DOI10.1109/3M-NANO.2016.7824965
会议(录)名称6th IEEE International Conference on Manipulation, Manufacturing and Measurement on the Nanoscale, IEEE 3M-NANO 2016
页码59-63
收录类别EI
会议地点Chongqing, China
会议日期July 18, 2016 - July 22, 2016